Lagus Applied Technology, Inc. is actively engaged in the development and use of tracer gas technology for a wide variety of measurement requirements. Consequently accessories, supplies and applications kits have been developed for use by purchasers of the AUTOTRAC®.
The electronic flowmeter is powered by an independent AC power module supplied with the accessory. The flowmeter incorporates a 3 1/2 digit LCD display. Each flowmeter is calibrated to within 2% of reading over a range of 20% to 100% of full scale. (For some ranges 1% FS is available at extra cost). Calibration data are supplied with the unit. Repeatability is 0.5% of full scale. Operational range is 0 to 50 C and 150 psi gauge (10 kg/cm2).
An electronics interface and appropriate software are installed in AUTOTRAC® to automatically read the electronic flowmeter and correct the reading for injection gas composition. The flowmeter is normally supplied with 1/4” tube fittings. Other fitting sizes can be supplied on request, including metric sizes.
Note to international clients: This accessory is intended for use by North American clients. Some authorized LAT distributors provide versions of this kit for their service area. Use of US approved gas cylinders and regulators may be illegal in your country and incompatible with locally available gas cylinders. Consult LAT or the distributor before ordering this kit.
Note to international clients: This accessory is intended for use by North American clients. Some authorized LAT distributors provide versions of this kit for their service area. Using US approved gas containers and regulators may be illegal in your country and incompatible with locally available gas cylinders. Consult LAT or the distributor before ordering this kit.
The module consists of five electronic mass flow meters, five injection control solenoid valves, five manual flow control metering valves and electronic interface for sensing five status events and 26 analog channels. These channels can be used to log pressure temperature, wind speed, etc. inputs from either LAT supplied or user supplied signals.
Each of the five injection channels incorporates separate supply and outlet connectors (1/4” OD tubing). The five event status indicators are designed to be used with flow indicator switches to insure that AUTOTRAC® injects tracer gas only when the HVAC systems are operating or when there is flow in the stacks or ducts. This feature can be disabled by the user.
Individual flow meters can be ordered with a full scale air flow reading that ranges from 10 cc/min to 40 liter/min. Note that the five flow meters need not possess the same full scale range. Flowmeter readings are accurate to 2 % FS. (For some ranges 1% FS is available at extra cost) Calibration data are supplied with the unit. Repeatability is 0.5% of full scale. Operational range is 0 to 50 C and 150 psi gauge (10 kg/cm2). An electronic interface and appropriate software are installed in AUTOTRAC® to automatically read each flow meter and to correct the reading for injection gas composition.
The injection control solenoid valves are bubble tight (leakage of less than 1x10-5 cc/sec helium) two way normally closed valves. The maximum inlet pressure is 50 psig with a Cv factor of 0.03. (Limited other pressure and Cv values can be supplied).
The 8-port Automatic Sampling Module should be installed to fully utilize this accessory.
The solenoid valve is a bubble tight (leakage of less than 1 x 10-5 cc/sec of helium) two way normally closed valve. It is usually supplied with 1/4” tube fittings. Other fitting types can be provided. Maximum inlet pressure is 50 psig with a Cv factor of 0.03. (Limited other pressure and Cv values can be supplied).